Test Principle
W401 2.0 water vapor transmission rate tester adopts the principle of infrared method. Fix the pre-treated sample in the middle of the test chamber, divide the test chamber into a high-humidity side and a low-humidity side, the humidified nitrogen flows on one side of the film, and the dry nitrogen (carrier gas) flows at a fixed flow rate on the other side, in the presence of a humidity gradient, water vapor will permeate from the high-humidity side to the low-humidity side, and the water vapor that passes through the sample is carried to the infrared sensor by the flowing dry nitrogen, and the water vapor transmission rate (and other parameters) of the sample is obtained by the electrical signal output by the sensor.
Working principle diagram
Standard
ASTM F1249, BS EN ISO 15106-2, JIS K7129, GB/T 26253, YBB 00092003
Specification
| Item | Technical Parameters |
|---|---|
| Testing Range | 0.002 - 100 g/(m²·24h) |
| Repeatability | Take the larger value between 0.005 and 2% |
| Resolution | 0.0001 g/(m²·24h) |
| Temperature Control Range | 15 - 45℃ |
| Temperature Control Precision | ±0.1℃ |
| Humidity Control Range | (5 - 90)%RH, 100%RH |
| Humidity Control Precision | ±1%RH |
| Permeation Area | 50.24 cm² |
| Specimen Size | Φ100 mm |
| Specimen Thickness | ≤3 mm |
| Number of Specimens | 1 piece |
| Carrier Gas | 99.999% Nitrogen (Provided by the User) |
| Carrier Gas Pressure | ≥0.1 MPa |
| Carrier Gas Flow Rate | 5 - 100 mL/min |
| Pneumatic Pressure | ≥0.3 MPa |
| Overall Dimensions | 380*620*385mm |
| Weight | 25 kg |
| Power | 450 W |
| Power Supply | AC 220 V, 50 Hz |
Features
| Patented core technology
High-precision original imported infrared sensor, high sensitivity, with ultra-high stability and ultra-low failure rate, resolution of 0.0001 g/(m2·24h). New pneumatic control system, automatic fixture one-button lock sample, convenient and labor-saving, superior sealing performance. |
Precise control
Temperature: The semiconductor stabilizer automatically controls the temperature, precision is 0.1℃. Humidity: Dual airflow (dry and wet) humidity control, high precision to ±1%RH. |
| Real-time visualization
Embedded 11.6-inch high-res LCD. Real-time display of transmission, temperature, humidity, flow, and voltage curves. |
Global certification
Intelligent system designed according to GMP "Computerized System" for full data traceability and audit tracking. |
| Intelligent IoT platform
Can be connected to the IoT platform for digital network management, remote diagnosis, and troubleshooting. |
|
Application
![]() |
Film | Water vapor transmittance test of various plastic films (PP/PET/PE/PVC/BOPP/CPP, etc.), composite films, paper-plastic, aluminized, and degradable films (PLA/PBAT/PBS, etc.). |
![]() |
Sheet | Water vapor transmittance test of solid pharmaceutical hard sheets (PP/PVC/PTP, etc.), metal composite sheets, rubber sheets and other flakes. |
![]() |
Cardboard | Water vapor transmittance test of coated paper, silicone paper, cigarette bag aluminized paper, and paper aluminum-plastic composite sheets. |
![]() |
Plasters | Water vapor transmission performance test of medical plasters. |
![]() |
Package | Customizable fixtures for pharmaceutical bottles, sealed bags, ointment tubes, infusion hoses, plastic trays, etc. |
Why Choose Us
111 (Sub-headline placeholder)
111 (Title 1)
111 (Description text for the first reason why customers should choose your technology or service.)
111 (Title 2)
111 (Description text for the second reason, such as professional R&D team or state-level high-tech enterprise status.)
111 (Title 3)
111 (Description text for the third reason, such as global certification, 24/7 service, or customized solutions.)