W203 2.0 Water Vapor Transmission Rate Tester
Electrolytic Method | High Barrier Materials WVTR Testing

Product Introduction
W203 2.0 is developed based on the electrolytic test principle, designed and manufactured in accordance with GB/T 21529, ISO 15106-3 and other international standards. It tests water vapor transmission rate of samples under controlled temperature and humidity, widely used for high barrier materials in LCD, solar cells, chip packaging, pharmaceuticals, food and more.
Test Principle
W203 2.0 adopts the electrolytic method. The pre-treated sample is fixed in the test chamber, separating it into wet and dry sides. Humidified nitrogen flows on one side of the film, while dry nitrogen (carrier gas) flows on the other side at a fixed rate. Due to humidity gradient, water vapor permeates from the wet side to the dry side, carried by dry nitrogen to the electrolytic sensor. The WVTR value is calculated from the electrical signal output by the sensor.
Compliant Standards
Technical Specifications
Key Features
- High-sensitivity sensor for ultra-high barrier materials in electronics, photovoltaics, displays.
- Automatic pneumatic clamping system, one-key locking, excellent sealing.
- Precise temperature & humidity control: ±0.1℃, ±1%RH.
- 3 independent test chambers for high-throughput testing.
- 11.6-inch HD touch screen, real-time test curves display.
- Intelligent OS with audit trail, multi-level user rights, GMP compliant.
- Calibrated with national standard substances for accurate & authoritative data.
Applications
LCD sealant, SiNx passivation film, modified PET, etc.
FR-4, BT resin, PI substrate, ABF film, ceramic substrate.
Solar backsheet, butyl film, ALD Al₂O₃, EVA/POE film.
Pharmaceutical tubes, BFS bottles, infusion hoses, containers.
Ultra-high barrier films and packages for medical and food use.